Overview
The Woodpecker Dpex V is an electronic apex locator that measures the root canal length to the apical foramen with multiple-frequency impedance measurement and automatic calibration. It is the Dpex V housing of the Woodpex V (same features and instructions) and can be connected to a compatible Woodpecker endo motor.
Uses
- Measuring canal length in pulpitis, pulp necrosis and periapical periodontitis
- Measuring tooth length before a post crown
- Measuring tooth length for transplantation and replantation
Key features
- New generation, more than accurate
- 25 national patents and 8 invention patents
- Newly upgraded algorithm: more stable, with stronger anti-interference
- Multi-frequency technology measures accurately even with blood or residual pulp in the canal; results not affected by tooth type; more accurate with rubber dam
- Colour 4.5" LCD shows the file position; buzzer alert as the file nears the apex
- Optional connection to a compatible Woodpecker endo motor for apex locating during preparation
- Tactile detection reaches only 42% accuracy within ±0.5 mm; X-ray shows 3D anatomy in 2D, and over 80% of apical foramina lie on the side of the root
Specifications
| Brand |
Woodpecker |
| Manufacturer |
Guilin Woodpecker Medical Instrument Co., Ltd., Guilin, Guangxi, China |
| Model |
DPEX V (Woodpex V in another housing) |
| Display |
4.5" colour LCD |
| Battery |
Rechargeable lithium 14500, 3.6 V / 750 mAh |
| Power supply |
Adapter input 100–240 V~ 50/60 Hz, output DC 5 V / 1 A; consumption ≤ 0.5 W |
| Dimensions |
138 × 77 × 107 mm |
| Weight |
385 g |
| Regulatory |
Class II; IPX0; continuous operation; type B applied parts (touch probe, lip hook, file clip) |
Package contents
- Main unit
- Measuring wire
- Lip hooks, file clip and touch probe
- Adapter and battery
- Tester
Storage, handling and shelf life
- At "0.0 / APEX" the file is at the anatomical apical foramen; working length = measured length minus 0.5–1 mm. The 0.5 / 1.0 / 2.0 marks are not distances in mm
- Endo files should not be used more than 3 times with the measuring clip
- Dry the access cavity before measuring; use a file size suited to the canal; keep the file and file clip away from other metal; take care with metal crowns or bridges
- Check a suspiciously short reading with an X-ray
- Use only the original adapter, lithium battery and accessories; remove the battery for long storage
- Do not use on patients (or by dentists) with a pacemaker or other implanted electrical equipment
- File clip, lip hook and touch probe: clean, then steam-sterilize before each use (at least 4 min at 132 / 134 °C, 2.0–2.3 bar; max 20 min at 134 °C); max 200 cycles for the file clip, 1,000 for the lip hook and touch probe
- Main unit and measuring wire: wipe with disinfectant
- Store the device at -20 to +55 °C, 10–93% RH; use at +5 to +40 °C, 30–75% RH; use sterilized accessories within 7 days or reprocess them