Overview
Prima Classic inverted cone burs are tungsten carbide operative burs from Prima Dental for use in a dental rotary handpiece, here in FG and RA shanks. The table below gives each bur's US number, ISO number, head diameter and shank.
Uses
- Restorative procedures with a dental rotary handpiece: cutting and grinding tooth and bone, and dental prosthesis materials (metal, plastic, porcelain); transient use (under 60 minutes)
- Recommended speed by application (IFU Table 1): cavity preparation in enamel or dentine, heads 010–023, below 450,000 rpm; removal of amalgam or composite fillings, heads 010–018, 60,000–120,000 rpm; excavation in enamel, dentine or bone, heads 010–023, below 2,000 rpm; cutting bone, heads 018–027, 500–3,000 rpm; crown and bridge metal finishing, heads 018–027, below 30,000 rpm; trimming prosthetic polymers, heads 018–027, below 20,000 rpm
Key features
- Brazed in the neck rather than at the base of the head: the 100% carbide neck delivers optimum concentricity and strength
- Engineered to cut more effectively, with greater blade contact on the tooth surface and reduced risk of chipping or breakage
- Fast, smooth, vibration-free performance
- Stainless steel shank: every bur can be autoclaved multiple times without the risk of corrosion
Specifications
| Brand |
Prima Dental |
| Manufacturer |
Prima Dental Manufacturing Ltd, Gloucester, United Kingdom |
| Range |
Prima Classic tungsten carbide operative burs |
| Shape |
Inverted Cone, Fig. 2 |
| Shanks |
FG friction grip, 19 mm overall; RA latch (contra-angle), 22 mm overall |
| Material |
Tungsten carbide with 10% cobalt, single piece or a carbide tip brazed with nickel to a surgical-grade stainless steel stem |
| Supplied |
Non-sterile; reusable after reprocessing |
| CE marking |
CE 0123, as printed on the instructions for use (non-sterile) |
Available options
| Option |
Matest code |
Maker reference |
Head Ø (mm) |
Shank |
| FG (19 mm) / 34- ISO 500 314 010 001 008 |
FG34 |
US No. 34; ISO 500 314 010 001 008 |
0.8 |
FG, 19 mm |
| FG (19 mm) / 35- ISO 500 314 010 001 010 |
FG35 |
US No. 35; ISO 500 314 010 001 010 |
1.0 |
FG, 19 mm |
| FG (19 mm) / 36- ISO 500 314 010 001 012 |
FG36 |
US No. 36; ISO 500 314 010 001 012 |
1.2 |
FG, 19 mm |
| FG (19 mm) / 37- ISO 500 314 010 001 014 |
FG37 |
US No. 37; ISO 500 314 010 001 014 |
1.4 |
FG, 19 mm |
| FG (19 mm) / 38- ISO 500 314 010 001 016 |
FG38 |
US No. 38; ISO 500 314 010 001 016 |
1.6 |
FG, 19 mm |
| RA (22 mm) / 35- ISO 500 204 010 001 010 |
RA35 |
US No. 35; ISO 500 204 010 001 010 |
1.0 |
RA, 22 mm |
| RA (22 mm) / 36- ISO 500 204 010 001 012 |
RA36 |
US No. 36; ISO 500 204 010 001 012 |
1.2 |
RA, 22 mm |
| RA (22 mm) / 37- ISO 500 204 010 001 014 |
RA37 |
US No. 37; ISO 500 204 010 001 014 |
1.4 |
RA, 22 mm |
| RA (22 mm) / 38- ISO 500 204 010 001 016 |
RA38 |
US No. 38; ISO 500 204 010 001 016 |
1.6 |
RA, 22 mm |
Package contents
Storage, handling and shelf life
- Supplied non-sterile: clean, disinfect and steam-sterilize before first use
- Validated steam cycle: pre-vacuum (4 pulses), 134 °C for at least 3 minutes, drying for at least 30 minutes; steam autoclave only
- Do not use chlorine or chloride cleaners; reprocess within 1 hour of use; do not store for long in disinfectant solutions
- Reusable after reprocessing; inspect before use and discard damaged burs
- In use: wear eye protection and a surgical mask; never exceed the maximum speed; no excessive pressure; seat the bur fully in the collet; irrigate properly, as too little irrigation can cause heat, discomfort, necrosis or burns; keep the bur moving
- Not recommended for people with an allergy or sensitivity to cobalt or, for nickel-brazed burs, to nickel
- Store dry and clean at ambient temperature, in the sterilization container (tray, bur stand or pouch) until needed
- Used burs are biohazard waste: discard them under local regulations
Documents
- Prima instructions for use IFU-27, carbide rotary dental burs, non-sterile, Revision A18 (2024-09-17)
- Prima Dental catalogue 2025 (2025-08-28), pages 3 and 4